Abstract

BACKGROUND. Tyrosine Kinase Inhibitors (TKIs) are part of the successful clinical management of patients with Chronic Myeloid Leukaemia (CML). However, optimal clinical management of CML requires a robust, standardized laboratory assay used at key clinical milestones to ensure a successful outcome for patients on TKIs. Quantitative monitoring of %BCR-ABL1IS by reverse transcription quantitative PCR (RT-qPCR) is the gold standard strategy for evaluating patient response to therapy and classification into prognostic subgroups. However, it can be challenging to perform in a reproducible manner. Reverse-Transcription Digital PCR (RT-dPCR) is an adaptation of this method that could provide the robust and standardized workflow needed for patient stratification. AIM. In this study, we compared three different dPCR platforms and investigated whether they could be applied to a clinical setting to quantify BCR-ABL1 transcripts in CML patients. METHODS. BCR-ABL1 and ABL1 transcript copy numbers were quantified in a total of 102 samples; 70 CML patients undergoing TKI therapy and 32 non-CML individuals. Three commercially available RT-dPCR platforms (QS3D, QX200 and RainDrop) were compared with the routinely used RT-qPCR platform using a modified version of the Europe Against Cancer (E.A.C.) assay. RESULTS. Measurements on all instruments correlated well when the %BCR-ABL1IS was ≥0.1% (R2 = 0.91, 0.93, 0.95 for QS3D, QX200, RainDrop, respectively)(Figure 1A). RT-dPCR quantification of the BCR-ABL1 transcript copy numbers however correlated well with RT-qPCR across all three platforms down to ≥0.1% (R2 = 0.85, 0.94, 0.92 for QS3D, QX200, RainDrop, respectively) (Figure 1B). Below this level, while the correlation was maintained, there was greater variation among instruments between RT-dPCR and RT-qPCR. Agreement between instruments was consistently higher for BCR-ABL1 copies than for ABL1 copies on the same patient set; thus changing the %BCR-ABL1 ratio values frequently varying by one order of magnitude when compared to RT-qPCR (R2 = 0.89, 0.92, 0.97 for QS3D, QX200, RainDrop, respectively) (Figure 1C). Furthermore, none of the platforms were able to substantially improve the sensitivity of quantification by RT-qPCR in relation to patients samples with %BCR-ABL1IS level When using the QS3D there was good correlation down to approximately 20 copies of BCR-ABL1 transcripts, below which a plateau was observed (Figure 2). For quantification of the lowest two CML categories, 0.001% and CONCLUSION. RT-dPCR was able to quantify low level BCR-ABL1 transcript copies but was unable to improve sensitivity below the level of detection achieved by RT-qPCR. However, RT-dPCR was able to perform these sensitive measurements without calibration curve. Adaptions to the protocol to increase the amount of RNA measured, a redesign of the assay used and an alternative reference gene to ABL1 are likely to be necessary to improve the analytical sensitivity of BCR-ABL testing. Disclosures Griffiths:Affymetrix: Research Funding. Apperley:Bristol Myers Squibb: Honoraria, Speakers Bureau; Incyte: Speakers Bureau; Pfizer: Honoraria, Speakers Bureau; Novartis: Honoraria, Speakers Bureau; Ariad: Honoraria, Speakers Bureau.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call