Abstract

An isocratic high-performance liquid chromatographic method was developed and validated for the simultaneous determination of human serum albumin (HSA) and timolol in albumin nanoparticles. This method involved a reversed-phase-C18 column thermostated at 25°C, UV detection at 276nm, flow rate of 1.0ml/min and a mobile phase compounded by 0.05% (v/v) trifluoroacetic acid in water/0.05% (v/v) trifluoroacetic acid in an acetonitrile (40:60, v/v) solution. The elution times for albumin and timolol were 1.84±0.05min and 2.67±0.04min, respectively. Calibration curves were linear from 0.2 to 100mg/ml for HSA and 0.01 to 1mg/ml for timolol. Limits of quantification were 0.2mg/ml for HSA and 0.01mg/ml for timolol. The values of accuracy and precision of intra- and inter-day variation studies were within acceptable limits, according to the US Food and Drug Administration Guidance for Industry. The described method has proved to be useful to give accurate measurements of human serum albumin and timolol from albumin nanoparticles to determine the percentage of encapsulation and the process yield.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.