Abstract
A new, simple, accurate, precise, robust and isocratic RP-HPLC method has been developed and subsequently validated for the determination of Regorafenib in pure form and pharmaceutical dosage forms as per ICH guidelines. The separation achieved on a Symmetry C18 Column, 250 mmx4.6 mm i.d. and 5μm particle size column as a stationary phase and Methanol: Phosphate buffer (pH adjusted to 4.80 with phosphoric acid) in the ratio of 70: 30v/v used as mobile phase at a flow rate of 1.0 ml/min. The UV detection was performed at 268nm. The retention time for Regorafenib was found to be 3.544minutes. The detector response was linear in the concentration range of 0–16μg/ml. The respective linear regression equation being Y=58945.x+9634 with R2=0.999. The percentage of Regorafenib in pharmaceutical dosage form was found to be within the limits. The limit of detection and the limit of quantification were found to be 0.90μg/ml and 2.90μg/ml respectively. The results of the study showed that, the proposed RP-HPLC method was simple, rapid, precise, accurate and stability indicating, which can be used for the routine determination of Regorafenib in pure form and pharmaceutical dosage forms.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.