Abstract

A simple, precise, accurate, economical and reproducible HPLC method for estimation of montelukast in tablet dosage form has been developed. Quantitative HPLC was performed with Shimadzu LC2010c HT with Winchrom Software with UV‐Visible detector (SPD‐IOA), PUMP (LC‐IOAT) and (LC‐IOATvp). Phenomenex C8, 5 μm, 25 cm × 4.6 mm i.d. column was used in the study. The mobile phase of ACN: Acetate buffer= 6.5:3.5 of pH 3 were used in this study. The conditions optimized were: flow rate (1 mL/minute), wavelength (222 nm) and run time was 20 min. Retention time was found to be 3.08 min. The linearity was found to be in the concentration range of 10‐100 μg/mL. The developed method was evaluated in the assay of commercially available tablet moni containing 10 mg of montelukast. The amount of drug in tablet was found to be 10.34 mg/tab for the brand. Results of analysis were validated statistically and by recovery studies. The recovery studies 99.67% was indicative of the accuracy of proposed method. The precision was calculated as repeatability, inter and intraday variation (%RSD) for the drug. By using the method, stability of the drug has been studied.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.