Abstract

Conventional monitoring of hemodialysis dose is implemented using urea kinetic modeling based on single-pool Kt/V, which requires both pre- and postdialysis serum urea nitrogen (SUN) measurements. We compared this conventional approach to one in which Kt/V is calculated using conductivity clearance, thereby reducing the need for regular postdialysis SUN measurements. Comparative study of 2 diagnostic tests. Prevalent patients receiving maintenance hemodialysis for at least 2 years for whom both urea reduction ratio (URR) and average conductivity clearance (Kecn) were measured. During the initial 8 months (baseline interval), average Kecn and URR were used to calculate a median patient-specific, modeled, calibration solute distribution volume (Vcal). During months 9 to 16 (period 1) and 17 to 24 (period 2), Kt/V was conventionally computed using URR and also by a new method using Vcal and Kecn without postdialysis SUN values. We examined the percentage error between these 2 methods of calculating Kt/V. Concordance between the 2 methods of calculating Kt/V. Among 1,093 patients, mean individual-level median single-pool Kt/V values derived using the conventional method during the baseline interval, period 1, and period 2 were 1.62±0.24 (SD), 1.66±0.24, and 1.67±0.24, respectively. During periods 1 and 2, patient-level median Kt/V values derived using Kecn were 1.64±0.24 and 1.65±0.24, respectively. Percent differences between patient-level median values of Kt/V (conductivity minus conventional URR methods) were-0.63%±7.7% and-0.75%±8.4% for periods 1 and 2. Normalized protein nitrogen appearance were comparable between the 2 methods. Data were collected over 2 years. Study was limited to in-center hemodialysis patients dialyzed 3 times per week. Dialysis session length was not adjusted for treatment interruptions. A new method of calculating Kt/V based on Kecn that requires fewer postdialysis SUN measurements provided diagnostic data comparable to those from conventional use of URR and has the potential to avoid errors related to postdialysis blood sampling and measurement.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call