Abstract

An analytical method has been developed using a spark-source mass spectrograph for rapid, routine analysis. Using an electrical detection system and reproducible sample positioning, precisions of ±3% are attained under routine conditions with an accuracy of ±7% for several elements in the NBS 460 series of iron standards. Repetitive analyses of Ag in Cu demonstrate the method can analyze trace elements in metals with accuracies approaching ±2%. Details of the electrode positioning procedure and the analytical accuracy and precision of the technique are presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.