Abstract
The round robin study of reference standards of 35 Cl , 32 S and 12 C implanted into Cu films using five SIMS instruments from Taiwan SIMS users was attempted. Significant differences in RSFs but similarity in maximum peak concentration among different instruments was observed. For 35 Cl , 32 S and 12 C measured using Cameca ims-6f, the RSF is 3.67E+19, 5.44E+20, 4.24E+21, the detection limit is 2.5E+16, 3.17E+16, 2.34E+18, and the dynamic range is 8.11E+3, 2.22E+4, 2.3E+1, respectively. The exchangeable use of RSF is recommended among model-specific instruments.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have