Abstract

We have designed a low-loss dielectric sample holder for positioning very thin, metallic, semiconductor samples in a TE102 cavity for ESR studies at liquid-helium temperatures. The sample holder can be rotated ±15° from the center position, allowing accurate positioning of the high-conductivity sample at minimum electric field. This results in maximizing the Q factor and allowing symmetric excitation on both faces of the sample. We have also devised a method of setting the microwave phase to desired value to better than 0.3° by utilizing the hyperfine lines of a dilute As-doped Si sample.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.