Abstract
Surface reflections often present a problem in the polarization difference imaging of tissue. The technique described involves illumination in a single polarization state and detection in the orthogonal polarization state. Synchronously rotating both the illumination and orthogonal detection states provides an image free from surface reflections that is sensitive to the polarization properties of the underlying tissue. Results on tissue phantoms demonstrate that polarization sensitive measurements of a test target can be made up to a depth of 12 mean free paths within a scattering medium. Preliminary images of bovine tendon are also demonstrated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.