Abstract

We present a volumetric sampling method that rotates the spiral interleaves of a stack of spirals (SOSP) trajectory for reduced aliasing artifacts using parallel imaging with undersampling. The aliasing pattern in an undersampled SOSP acquisition was modified by consecutively rotating spiral interleaves in each phase-encoding plane. This allows a sampling scheme with a high reduction factor when using a volumetric multireceiver array. Phantom and in vivo brain images at a resolution of 1 × 1 × 2 mm(3) were acquired at 3T using a 32-channel coil. Images reconstructed with a reduction factor of 16 were compared for aliasing artifacts and geometry factor (g-factor). Phantom and in vivo brain image results revealed that the rotated SOSP acquisition with a reduction factor of 16 produces images with reduced aliasing and lower g-factors than images acquired without rotation. The proposed rotated SOSP sampling method is a highly efficient way to maximize the encoding power of volumetric receiver arrays in parallel imaging and is applicable to rapid volumetric scanning, including susceptibility-weighted imaging and functional MRI. Magn Reson Med 76:127-135, 2016. © 2015 Wiley Periodicals, Inc.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.