Abstract
Prolonged exposure to elevated temperatures exceeding 47°C, which can occur during root canal obturation, can cause damage of both dental and bone tissues. In order to study the temperature distribution on the surface of the tooth root a temperature measuring device with cold-junction compensation is proposed. For in vitro measurement of the temperature distribution on the surface of the tooth, 8 thermocouples placed in direct contact with the cementum of the tooth were used. In order to eliminate the cold-junction temperature variations, the temperature equilibration device and RTD were used. The suggested linear approximation for the thermocouples' conversion function provides a nonlinearity relative error of less than 0.05% for K-type thermocouples and 0.07% for J-type thermocouples over the temperature range from 20 to 60°C.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Informatyka, Automatyka, Pomiary w Gospodarce i Ochronie Środowiska
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.