Abstract

A simple and sensitive method is described for the determination of biogenic indole compounds adsorbed on cellulose or alumina plates for thin-layer chromatography by room-temperature phosphorimetry. The optimum conditions were investigated for 5-hydroxyindole-3-acetic acid and indole-3-acetic acid. The compounds are spotted on the plates, which are then sprayed successively with sodium citrate or sodium acetate, and sodiu iodide solutions. The plates are dried completely under a stream of dry nitrogen, and immediately dipped in molten paraffin. The phosphorescence is stable for at least 3 h even in moist air. The limits of detection for nine biogenic indole compounds tested are between 2 and 300 pmol per sample spot.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.