Abstract

In this paper, the replication possibilities of microfluidic channels by UV-roll-to-plate fabrication were investigated and a study of rheology-modified thiol-ene for the application in such a UV-roll-to-plate setup was conducted. The system allows the manufacture of channels with aspect ratios of 2:1 and a maximal channel depth of 90 μm as well as the sealing of the finished devices with patterning and sealing speeds of up to 19 m min−1. By adding fumed silica nanoparticles to the uncured resins, it was possible to alter the rheological behavior of the resin system to fabricate shallow microfluidic channels with 40 × 95 μm cross-sectional dimensions. Moreover, deeper (90 μm) channels can be fabricated with highly viscous resins based on thiol-terminated oligomers. As a demonstration, capillary electrophoresis chips were prepared and tested for a simple separation of two fluorescent dyes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.