Abstract

In this work, we investigate the effect of interdigitated electrode fingers width and gap on the performance of biosensors. Two different configurations of interdigitated electrodes were utilized; one consists of 50 μm wide fingers with a gap of 50 μm, and the other consists of 100 μm wide fingers with a 100 μm gap. First, the active surface area of both interdigitated electrodes was investigated. Then, the surface electronic parameters were investigated using faradaic and non-faradaic electrochemical impedance spectroscopy modes. Both interdigitated electrodes were then functionalized with LDL-antibodies through a standard chemical functionalization route and were utilized for the detection of LDL-antigen of cholesterol. The biosensor with 50 μm width and gap demonstrated better sensitivity and limit of detection of 6.08 nF/log(ng/ml) and 40 pg ml−1, respectively. Biosensors with 100 μm finger width and gap sensitivity and detection limit are 5.16 nF/log(ng/ml) and 300 pg ml−1, respectively. The enhanced sensitivity and limit of detection for the first biosensor is due to the increased surface area, enhanced electric field, and less diffusion of ions onto the biosensor surface. The study reported here suggests room for improving the geometry of many interdigitated electrodes for enhanced non-faradaic biosensing performances.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.