Abstract

Percutaneous aortic valve replacement (PAVR) is an emerging therapy for nonsurgical patients with severe aortic stenosis (AS). We examined the role of transesophageal echocardiography (TEE) in PAVR. TEE was used initially to assess the native valve and aortic root, and served as a guide during PAVR. Following prosthetic valve deployment, TEE was used to assess valve function. Eleven patients aged 82 +/- 10 years with NYHA III-IV underwent PAVR. Periprocedural TEE gave immediate information on prosthetic position and function, LV function, mitral regurgitation, pericardium, and thoracic aorta anatomy. There was excellent visual agreement between fluoroscopic and TEE images of prosthetic positioning and deployment. TEE facilitated the detection and management of procedure-related complications. Compared with pre-PAVR, AV area (0.56 +/- 0.19 cm(2) vs. 1.3 +/- 0.4 cm(2); P < 0.001) and LVEF (49 +/- 17% vs. 56 +/- 11%; P < 0.001) increased. TEE provides key anatomical and functional information, and serves as a diagnostic guide for complications, which may arise during PAVR.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call