Abstract

We report the results of experimental studies on the effects of sample supports in laser desorption/ionization mass spectrometry (LDI-MS). LDI time-of-flight (TOF) mass spectra obtained for C(60) and insulin samples deposited onto standard stainless steel substrate and/or onto some non-metallic materials (glass, scotch tape, floppy disc foil, Teflon foil, photocopy film), all recorded under identical, typical experimental conditions, have been compared with regard to their intensity and quality. The LDI investigations show that compared with stainless steel, glass and floppy disc foil sample supports boost (2-3.5 times) ion yields for C(60)(+) and C(60)(-) ions, respectively. The stainless steel and scotch tape sample supports are the best for the mass resolution of positive ions and the formation of (C(60))(n)(-) (n <or= 4) cluster ions, respectively. In the case of detection of insulin by matrix-assisted laser desorption/ionization (MALDI) we did not observe significant differences in sensitivity for the support materials tested. A mechanism of ion formation in the desorption plume is suggested.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.