Abstract

Thermoluminescence (TL) in LiF was found to depend on the use of proper combination of the dopant ions and the optimised preparation procedure. LiF doped with Mg,Cu,Si exhibited high TL sensitivity (55 times that of TLD-100 LiF:Mg,Ti) and insignificant higher temperature glow peak leading to negligible residual TL signal. The occurrence of the dosimetric peak in LiF was found to be a strong function of Mg dopant. The peak temperature and the TL emission spectra of dosimetric glow peak in all samples having Mg, Cu were always found to be the same irrespective of the third dopant which influenced only the intensity of TL emission spectrum. The emission band peaking at 385 nm was assigned to Cu.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.