We present a new approach for the accurate reconstruction of three-dimensional skeletal positions using roentgen single-plane photogrammetric analysis (RSPA). This technique uses a minimum of three markers embedded in each segment which allow continuous, real-time, internal skeletal movement to be measured from single-plane images, provided that the precise distance between the markers is known. A simulation study indicated that the error propagation in this approach is influenced by focus position, object position, the number of control points, the accuracy of the previous measurement of the distance between markers and the accuracy of image measurement. For reconstruction of normal movement of the knee with an input measurement error of SD = 0.02 mm, the rotational and translational differences between reconstructed and original movement were less than 0.27 degrees and 0.9 mm, respectively. Our results showed that the accuracy of RSPA is sufficient for the analysis of most movement of joints. This approach can be applied in combination with force measurements for dynamic studies of the musculoskeletal system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.