Abstract

This was a laboratory investigation. Rod attachment can induce significant pedicle screw-and-rod pre- strain that may predispose the instrumentation to failure. This study investigated how in vitro L5-S1 rod strain and S1 screw strain during rod-screw attachment (pre-strain) compared with strains recorded during pure-moment bending ( test- strain). The lumbosacral junction is highly vulnerable to construct failure due to rod fatigue fracture, sacral screw pull-out, and screw fatigue fracture. Twelve cadaveric specimens were instrumented with L2-ilium pedicle screws and rod. Strain gauges on contoured rods and sacral screws recorded strains during sequential rod-to-screw tightening (pre-strains). The same instrumented constructs were immediately tested in a 6-degree-of-freedom apparatus under continuous loading to 7.5Nm in multidirectional bending while recording instrumentation test-strains. Rod and screw pre-strains and test-strains were compared using 1-way repeated-measures analysis of variance followed by Holm-Šidák paired analysis (significant at P <0.05). The mean first (171±192µE) and second (322±269µE) rod attachment pre-strains were comparable to mean test-strains during flexion (265±109µE) and extension (315±125µE, P ≥0.13). The mean rod attachment pre-strain was significantly greater than mean test-strains during bidirectional lateral bending (40±32µE ipsilateral and 39±32µE contralateral, P <0.001) and axial rotation (72±60µE ipsilateral and 60±57µE contralateral, P <0.001). The mean first and second sacral screw pre-strains during rod attachment (1.03±0.66 and 1.39±1.00Nm, respectively) did not differ significantly ( P =0.41); however, the mean sacral screw pre-strain during final (second) rod attachment was significantly greater than screw test-strains during all directions of movement (≤0.81Nm, P ≤0.03). Instrumentation pre-strains imposed during in vitro rod-screw attachment of seemingly well-contoured rods in L2-ilium fixation are comparable to, and at times greater than, strains experienced during in vitro bending. Spine surgeons should be aware of the biomechanical consequences of rod contouring and attachment on construct vulnerability.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.