Abstract

TOF (train-of-four) monitoring provides objective data in application of neuromuscular blocking agent. Thus, applicator-based differences are eliminated and optimum muscle relaxation is maintained during operation. In the present study, we aimed to compare the effects of target-controlled infusion system and standard TOF monitoring, on use of rocuronium. ASA I-II patients, who were aged between 18 and 75years and scheduled for elective abdominal surgery at Haydarpaşa Numune Training and Research Hospital, were enrolled in the study. In order to evaluate neuromuscular blockade, the patients in Group 1 were connected to the acceleromyography device of the target-controlled infusion pump (Veryark-CLMRIS-I-China) while the ones in Group 2 were connected to the routinely used acceleromyography device (TOF Watch SX). There was no significant difference between groups regarding patient characteristics, the durations of anaesthesia and surgery, quality of intubation, time to extubation and time to recovery (TOF ratio of 0.9). Intubation time was significantly longer in Group 1 (Automated group) as compared to Group 2 (Control group) (p<0.05). The total rocuronium amount used in Group 1 was found to be significantly higher than the amount used in Group 2 (p<0.05). There was no clinical evidence of residual neuromuscular blockage or reoccurrence of neuromuscular blockage in any patient in either group. Both methods can be used for administration of neuromuscular blocker agent during moderate time anesthesia. No advantage was noted when rocuronium was administered via automatical infusion pump during anaesthesia.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call