Abstract
Abstract Left ventricular assist devices (LVADs) have become a viable alternative to heart transplantation in heart failure therapy. In clinical practice, rotary blood pumps used as LVADs are operated at a constant rotational speed and thus do not adapt to the varying demand of the patient. This paper presents a robust control approach for automatic adaptation of the blood pump speed to the blood flow demand of the patient’s body, which enables a defined load sharing between an LVAD and the native ventricle. Robust stability was checked using a detailed model of the human cardiovascular system with uncertainties that describe the most important native physiological control loops as well as a range of pathologies. The robust assistance controller was tested in an in vivo setup and was able to stabilize the cardiovascular system after myocardial infarction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.