An efficient nuclear magnetic resonance (NMR)-based method was developed to detect and quantify the presence of phthalates in plastic fragments obtained from a children’s toy and cosmetic product containers. Bis-2-ethyl hexyl phthalate (BEHP) was present at a concentration of 8.5% w/w in the plastic toy residue. This indicates levels of phthalates above the permitted threshold (e.g., <0.1% w/w) imposed by EU and USA regulations restricting the use of plasticizers in plastic toys for children aged 0 to 3 years. An added value of this method is that the tetramethylsilane (TMS) internal standard present at a concentration of 2.2 mmol/L in commercially available CDCl3 solvents may be conveniently used for the quantitative nuclear magnetic resonance (qNMR) analysis. The protocol described here is highly reproducible and may be easily applied to large scale investigations of plasticizers content into polyvinyl chloride (PVC) polymers.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.