Abstract

Arterial oxygen saturation (SpO₂) monitoring for newborns requires special attention in neonatal intensive care units (NICUs). Newborns have very low photo-plethysmogram (PPG) amplitudes and their body movements are difficult to contain. Hardware design and its associated signal processing algorithms should be robust enough so that faulty measurements can be avoided. In this study, improved designs were implemented to deal with low perfusion, motion artifact, and the influence of ambient light. Dynamic range was increased by using different LED intensities and a feedback system. To minimize the effects of motion artifact and to discard other unqualified data, four additional algorithms were used, which were based on dual-trace detection, continuity of DC level, morphology of PPG, and simultaneity check of SpO₂. Our SpO₂ system was tested with newborns with normal respiration in the NICU. Our system provided fast, real-time responses and 100% artifact detection was accomplished under 84% of SpO₂.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.