Abstract

This paper reports on the development of a new closed-bore magnetic resonance imaging (MRI) compatible robotic system for image-guided prostatic interventions: ablation, brachytherapy, and biopsy. The first stage of development addresses only laser-based ablation. The robot actuators are ultrasonic motors. The first physical robot prototype was manufactured and tested in the MRI with an ablation tool. The tests covered magnetic resonance (MR) compatibility, tool visualization, and robot control accuracy. Robot tip position error is less than 2 mm at points closer than 0.5 m to the isocenter. A method to control ultrasonic motors for MRI-compatibility is reported.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.