Abstract
BackgroundStereotactic irradiation of large or critically located arteriovenous malformations (AVMs) is a special challenge for clinicians and radiation physicists. To date, no comprehensive comparison of two linac-based radiosurgery systems used for hypofractionated radiotherapy of large AVMs was published. The aim of the study was to compare dose distributions between CyberKnife (CK) system and linac with a micro-multileaf collimator (L-mMLC) in high-grade or critically located cerebral AVMs.MethodsTwo sets of plans made for 15 different patients with at least 95% target coverage were selected for comparisons. Conformity (CI), homogeneity (HI) and gradient score (GSI) indices, conformity index proposed by Lomax (CIL), conformation number (CN), quality of coverage (Q), volumes of brain receiving 12,10,8,6,4, and 2 Gy, minimum and maximum doses for critical structures in both treatment planning systems (TPS) were compared. Finally, the number of monitor units needed to deliver the prescribed dose was compared.ResultsThe mean minimum doses in the target volume were 93.3% (CK) and 90.7% (L-mMLC),p=n.s, maximum: 119.7 and 110%, respectively (p=0.004). The mean CI was 1.46 and 1.86, HI: 1.2, and 1.11, CIL 0.7, and 0.6, CN: 0.68 and 0.58 for CK and mMLC, respectively (p<0.05). The values of GSI and Q were not significantly different. The volumes of the brain receiving low doses (4 Gy and 2 Gy) were significantly lower in the CK system. The number of monitor units necessary to deliver the prescribed dose was significantly greater in case of the CK system.ConclusionsBetter conformity can favor the CK system for treatment of large AVMs at the cost of higher maximum doses and worse homogeneity. L-mMLC is superior when shorter treatment time is required. Neither system can assure satisfying dose gradients outside large targets surrounded by numerous critical structures.
Highlights
Stereotactic irradiation of large or critically located arteriovenous malformations (AVMs) is a special challenge for clinicians and radiation physicists
In three plans made for linac with a micro-multileaf collimator (L-micro-multileaf collimator (mMLC)) and in one for the CK system, the dose constraints for critical structures were violated
The minimum doses for the target volumes were comparable, whereas the maximum doses were significantly larger in the CK system than in case of L-mMLC, p
Summary
Stereotactic irradiation of large or critically located arteriovenous malformations (AVMs) is a special challenge for clinicians and radiation physicists. No comprehensive comparison of two linac-based radiosurgery systems used for hypofractionated radiotherapy of large AVMs was published. The aim of the study was to compare dose distributions between CyberKnife (CK) system and linac with a micro-multileaf collimator (L-mMLC) in high-grade or critically located cerebral AVMs. Large and critically located arteriovenous malformations are consistently a challenge for clinicians. A large populational study conducted by the Finnish group provided evidence that active, even partial treatment may be of benefit for patients with cerebral AVMs [2]. AVMs involving critical structures can be treated with hypofractionated stereotactic radiotherapy (HFSRT), whereas lesions of large volume are suitable both to volume-staged radiosurgery and HFSRT [3,4]
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.