Abstract

We are developing an image-guided robot-based system to assist orthopaedic surgeons in performing distal locking of long bone intramedullary nails. The system consists of a bone-mounted miniature robot fitted with a drill guide that provides rigid mechanical guidance for hand-held drilling of the distal screws’ pilot holes. The robot is automatically positioned so that the drill guide and nail distal locking axes coincide using a single fronto-parallel fluoroscopic X-ray. This paper describes new methods for accurate and robust drill guide and nail hole localization and registration and reports the results of our in-vitro system accuracy experiments. Tests of 17 runs show a mean angular error of 1.3 o (std = 0.4 o ) between the computed drill guide axes and the actual locking holes axes, and a mean 3.0mm error (std = 1.1mm) in the entry and exit drill point, which is adequate for successfully locking the nail.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.