Abstract

PurposeTo provide a technical description of robot-assisted uterine artery embolization and to investigate the safety and feasibility of the Magellan (Hansen Medical, Mountain View, California, USA) robotic catheter in this complex arterial bed. Materials and MethodsFive women (mean age, 48.8 y) underwent robot-assisted bilateral uterine artery embolization over a 10-month period using the Magellan robotic catheter. Demographic, clinicopathologic, and endovascular performance metric data (fluoroscopy and cannulation times) were recorded as well as short-term outcomes. ResultsRobotic cannulation of bilateral internal iliac and uterine arteries was successful in all cases. Median right and left internal iliac artery cannulation and total fluoroscopy times were 3 minutes (interquartile range [IQR], 1.5–4 min), 2 minutes (IQR, 1.5–4 min), and 11 minutes (IQR, 9.5–14 min). Median right and left uterine artery cannulation times were both 11 minutes (IQR, 6.5–15 min and 8–12 min, respectively). Technical success was 100%. All patients were discharged on postoperative day 1, and there were no major or access site complications. At 6 months after the procedure, all patients reported significant improvement of symptoms, with a median increase in health-related quality-of-life score of 58% (48.5%–61.75%). ConclusionsThe use of the new-generation Magellan system in uterine artery embolization is feasible and appears to be safe. The additional navigational capability and added maneuverability of the NorthStar catheter (Hansen Medical, Mountain View, California) may facilitate selective catheterization of small iliac artery divisions and may be useful in any procedure where complex arterial selection is needed.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call