Abstract

A number of stereotactic platforms are available for performing deep brain stimulation (DBS) lead implantation. Robot-assisted stereotaxy has emerged more recently demonstrating comparable accuracy and shorter operating room times compared with conventional frame-based systems. To compare the accuracy of our streamlined robotic DBS workflow with data in the literature from frame-based and frameless systems. We retrospectively reviewed 126 consecutive DBS lead placement procedures using a robotic stereotactic platform. Indications included Parkinson disease (n = 94), essential tremor (n = 21), obsessive compulsive disorder (n = 7), and dystonia (n = 4). Procedures were performed using a stereotactic frame for fixation and the frame pins as skull fiducials for robot registration. We used intraoperative fluoroscopic computed tomography for registration and postplacement verification. The mean radial error for the target point was 1.06 mm (SD: 0.55 mm, range 0.04-2.80 mm) on intraoperative fluoroscopic computed tomography. The mean operative time for an asleep, bilateral implant without implantable pulse generator placement was 238 minutes (SD: 52 minutes), and skin-to-skin procedure time was 116 minutes (SD: 42 minutes). We describe a streamlined workflow for DBS lead placement using robot-assisted stereotaxy with a comparable accuracy profile. Obviating the need for checking and switching coordinates, as is standard for frame-based DBS, also reduces the chance for human error and facilitates training.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call