Abstract

The use of transcutaneous electrical nerve stimulation (TENS) for pain relief is increasing. At the same time the implantable cardioverter defibrillator (ICD) is a routine treatment for malignant tachyarrhythmias. Today patients often need devices for more than one condition, and consideration must be given to the interaction between them. We studied the risk of interference between TENS and the ICD function. Thirty patients who had received an ICD underwent a test protocol including TENS at the mammilla and hip levels, at two energy levels, and at the highest comfortable stimulation level. The effects of TENS on the electrocardiogram lead II, intracardiac electrograms, and the ICD marker channels were analyzed. Disturbance from TENS on the sensing function was seen at all stimulation attempts. Interference between the systems was observed in 16 patients. In eight patients (27%) the interpretation was VT/VF and in 14 patients (47%) as ventricular premature extra beats. Other kinds of interactions were seen in five patients (16%). Each patient could have more than one kind of interference. Noise reversion and undersensing might prevent the ICD from delivering shock when it should and the interpretation as VT/VF could result in inappropriate shocks. Because of the potentially serious consequences of interference we do not recommend the use of TENS in patients with ICD.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.