Abstract

The rate of blood contamination of IV tubings used in anaesthesia practice was investigated. Only IV tubings started in the operating room were studied. First, 300 tubings of three different types were tested at the three distal injection sites. The contamination rate was 3.3 per cent at the injection site closest to the IV catheter and 0.3 per cent at the furthest. The presence of a check-valve did not affect the contamination rate. Second, 300 third injection sites fixed at a level equal to or above the IV catheter were tested. None of them was contaminated. Finally, in order to evaluate whether changing the needle alone could prevent the contamination of syringes, injections were made into a tubing where blood was flowing. Thirty-four per cent of the syringes tested positive for blood. We conclude that IV tubings have a significant contamination rate in usual practice. This rate decreases as the distance from the IV catheter increases. The use of the third site fixed at a level equal to or above the IV catheter carries a lower risk of contamination. Changing the needle alone is a useless procedure to prevent cross-contamination.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.