Abstract

Hemolysis during pediatric extracorporeal membrane oxygenation (ECMO) is associated with increased risk for renal failure and mortality. We aim to describe risk factors for hemolysis in pediatric ECMO supported by centrifugal pumps. We conducted an analysis of retrospective data collected at an academic children's hospital from January 2017 to December 2019. Plasma-free hemoglobin (PFH) levels were measured daily, and hemolysis was defined as PFH>50mg/dL. Of 46 ECMO runs over 528 ECMO days, hemolysis occurred in 23 (58%) patients over a total of 40 (8%) ECMO days. In multivariable logistic regression models, VA-ECMO (aOR=4.69, 95% CI: 1.01-21.83) and higher hemoglobin (aOR = 1.38, 95% CI: 1.06-1.81) were independently associated with hemolysis. There were also non-significant trends toward increased risk for hemolysis with higher rotational pump speed (aOR=2.39, 95% CI: 0.75-7.65), higher packed red blood cell transfusions (aOR=1.15, 95% CI: 0.99-1.34), and higher cryoprecipitate transfusions (aOR=2.01, 95% CI: 0.83-4.86). Isolated pump exchanges that were performed in 12 patients with hemolysis led to significant decreases in PFH levels within 24h (89 vs 11mg/dL, p<0.01). Hemolysis is common in pediatric ECMO using centrifugal pumps. Avoidance of high pump speeds and conservative administration of blood products may help to mitigate the risk for hemolysis. Furthermore, pump exchange may be an effective first-line treatment for hemolysis.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call