Abstract

BackgroundThe aims of this meta-analysis were to: (1) validate the outcome of modern dual mobility (DM) designs in patients who had undergone primary and revision total hip arthroplasty (THA) procedures and (2) to identify factors that affect the outcome.MethodsWe searched for studies that assessed the outcome of modern DM-THA in primary and revision procedures that were conducted between January, 2000 to August, 2020 on PubMed, MEDLINE, Cochrane Reviews and Embase. The pooled incidence of the most common failure modes and patient reported outcomes were evaluated in patients who have received: (1) primary THA, (2) revision THA for all causes or (3) for recurrent dislocation. A meta-regression analysis was performed for each parameter to determine the association with the outcome. The study design of each study was assessed for potential bias and flaws by using the quality assessment tool for case series studies.ResultsA total of 119 studies (N= 30016 DM-THAs) were included for analysis. The mean follow-up duration was 47.3 months. The overall implant failure rate was 4.2% (primary: 2.3%, revision for all causes: 5.5%, recurrent dislocation: 6.0%). The most common failure modes were aseptic loosening (primary: 0.9%, revision for all causes: 2.2%, recurrent dislocation: 2.4%), septic loosening (primary:0.8%, revision for all causes: 2.3%, recurrent dislocation: 2.5%), extra-articular dislocation (primary:0.6%, revision for all causes:1.3%, recurrent dislocation:2.5%), intra-prosthetic dislocation (primary:0.8%, revision for all causes:1.0%, recurrent dislocation:1.6%) and periprosthetic fracture (primary:0.9%, revision for all causes:0.9%, recurrent dislocation:1.3%). The multi-regression analysis identified younger age (β=-0.04, 95% CI -0.07 – -0.02) and female patients (β=3.34, 95% CI 0.91–5.78) were correlated with higher implant failure rate. Age, gender, posterolateral approach and body mass index (BMI) were not risk factors for extra-articular or intra-prosthetic dislocation in this cohort. The overall Harris hip score and Merle d’Aubigné score were 84.87 and 16.36, respectively. Level of evidence of this meta-analysis was IV.ConclusionModern dual-mobility designs provide satisfactory mid-term implant survival and clinical performance. Younger age and female patients might impact the outcome after DM-THA. Future research directions should focus on, (1) long-term outcome of modern dual-mobility design, including specific concerns such as intra-prosthetic dislocation and elevated metal ion, and (2) cost-effectiveness analysis of dual-mobility implant as an alternative to conventional THA for patients who are at high risk of dislocation.

Highlights

  • The aims of this meta-analysis were to: (1) validate the outcome of modern dual mobility (DM) designs in patients who had undergone primary and revision total hip arthroplasty (THA) procedures and (2) to identify factors that affect the outcome

  • The first generation DM design was associated with higher aseptic loosening and intra-prosthetic dislocation (IPD) rate, which resulted from polyethylene wear, suboptimal fixation and surface coating of the acetabular component [24,25,26,27,28,29,30]

  • In the late 1990’s, a newer DM design was introduced with several modifications including modular design, shape, surface coating and highly cross-linked polyethylene to reduce the rate of aseptic loosening and IPD [31,32,33,34]

Read more

Summary

Introduction

The aims of this meta-analysis were to: (1) validate the outcome of modern dual mobility (DM) designs in patients who had undergone primary and revision total hip arthroplasty (THA) procedures and (2) to identify factors that affect the outcome. The cause of a dislocated prosthesis can be multifactorial, including both surgeon and patient related factors [8,9,10,11,12,13,14,15,16,17,18]. The first generation DM design was associated with higher aseptic loosening and intra-prosthetic dislocation (IPD) rate, which resulted from polyethylene wear, suboptimal fixation and surface coating of the acetabular component [24,25,26,27,28,29,30]. In the late 1990’s, a newer DM design was introduced with several modifications including modular design, shape, surface coating and highly cross-linked polyethylene to reduce the rate of aseptic loosening and IPD [31,32,33,34]

Objectives
Methods
Results
Discussion
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.