Abstract

We prospectively investigated the complications associated with intraparenchymal intracranial pressure (ICP) monitoring using the Camino intracranial pressure device. A fiberoptic ICP monitoring transducer was implanted in 631 patients. About half of the patients ( n = 303) also received an external ventricular drainage set (EVDS). The durations (mean ± SD) of ICP monitoring in patients without and with an EVDS were 6.5 ± 4.4 and 7.3 ± 5.1 days, respectively. Infection occurred in 6 patients with only an ICP transducer (6/328, 1.8%) and 24 patients with an EVDS also (24/303, 7.9%). The duration of monitoring had no effect on infection, whereas the use of an EVDS for more than 9 days increased infection risk by 5.11 times. Other complications included transducer disconnection (2.37%), epidural hematoma (0.47%), contusion (0.47%), defective probe (0.31%), broken transducer (0.31%), dislocation of the fixation screw (0.15%), and intraparenchymal hematoma (0.15%). In conclusion, intraparenchymal ICP monitoring systems can be safely used in patients who either have, or are at risk of developing, increased ICP.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.