Abstract

BackgroundGuidelines and text-book descriptions of the Rinne test advise orienting the tuning fork tines in parallel with the longitudinal axis of the external auditory canal (EAC), presumably to maximise the amplitude of the air conducted sound signal at the ear. Whether the orientation of the tuning fork tines affects the amplitude of the sound signal at the ear in clinical practice has not been previously reported. The present study had two goals: determine if (1) there is clinician variability in tuning fork placement when presenting the air-conduction stimulus during the Rinne test; (2) the orientation of the tuning fork tines, parallel versus perpendicular to the EAC, affects the sound amplitude at the ear.MethodsTo assess the variability in performing the Rinne test, the Canadian Society of Otolaryngology – Head and Neck Surgery members were surveyed. The amplitudes of the sound delivered to the tympanic membrane with the activated tuning fork tines held in parallel, and perpendicular to, the longitudinal axis of the EAC were measured using a Knowles Electronics Mannequin for Acoustic Research (KEMAR) with the microphone of a sound level meter inserted in the pinna insert.Results47.4 and 44.8 % of 116 survey responders reported placing the fork parallel and perpendicular to the EAC respectively. The sound intensity (sound-pressure level) recorded at the tympanic membrane with the 512 Hz tuning fork tines in parallel with as opposed to perpendicular to the EAC was louder by 2.5 dB (95 % CI: 1.35, 3.65 dB; p < 0.0001) for the fundamental frequency (512 Hz), and by 4.94 dB (95 % CI: 3.10, 6.78 dB; p < 0.0001) and 3.70 dB (95 % CI: 1.62, 5.78 dB; p = .001) for the two harmonic (non-fundamental) frequencies (1 and 3.15 kHz), respectively. The 256 Hz tuning fork in parallel with the EAC as opposed to perpendicular to was louder by 0.83 dB (95 % CI: −0.26, 1.93 dB; p = 0.14) for the fundamental frequency (256 Hz), and by 4.28 dB (95 % CI: 2.65, 5.90 dB; p < 0.001) and 1.93 dB (95 % CI: 0.26, 3.61 dB; p = .02) for the two harmonic frequencies (500 and 4 kHz) respectively.ConclusionsClinicians vary in their orientation of the tuning fork tines in relation to the EAC when performing the Rinne test. Placement of the tuning fork tines in parallel as opposed to perpendicular to the EAC results in a higher sound amplitude at the level of the tympanic membrane.

Highlights

  • Guidelines and text-book descriptions of the Rinne test advise orienting the tuning fork tines in parallel with the longitudinal axis of the external auditory canal (EAC), presumably to maximise the amplitude of the air conducted sound signal at the ear

  • Audiology society recommendations [5] instructions aimed at medical student and non-specialist on performing the Rinne test in general and otolaryngology textbooks [6], instructions intended for otolaryngology residents in speciality textbooks [7], and peer reviewed

  • An experimental simulation of the air conduction component of the Rinne test was used to measure the sound intensity at the level of the tympanic membrane for both parallel and perpendicular positions of the tuning fork

Read more

Summary

Introduction

Guidelines and text-book descriptions of the Rinne test advise orienting the tuning fork tines in parallel with the longitudinal axis of the external auditory canal (EAC), presumably to maximise the amplitude of the air conducted sound signal at the ear. Mathematical calculations and sound field recordings conclude that a higher amplitude sound is delivered to the ear when the fork is placed parallel to as opposed to perpendicular to the EAC [9, 10]. These lines of evidence show a 5 dB difference in the sound intensity produced by the two different positions of the tuning fork [10]. The sound intensities of a tuning fork placed parallel to and perpendicular to the EAC during the Rinne test have not been compared before

Methods
Findings
Discussion
Conclusion