Abstract

The authors present two cases of patients with perforation of the right ventricular wall by the implantable cardioverter defibrillator (ICD) lead. The complication was resolved by cardiosurgical revision and epicardial leads stitched onto the diaphragmatic wall of the heart. The perforation was identified by electrical parameter changes of the leads, echocardiography, and computed tomography. Both patients had satisfactory values of electrical parameters and ICD function with epicardial leads. The importance of regular follow-up and a check of the lead parameters are emphasized.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.