Abstract
The adaptive cardiac resynchronization therapy (CRT) algorithm provides synchronized left ventricular pacing (sLVP). However, ensuring a high sLVP rate is challenging. We assessed the association between the sLVP rate and pacing sites in the right atrium. We evaluated 71 patients who underwent CRT and in whom the adaptive CRT algorithm was applied (53 men; mean age, 66 ± 14 years; median follow-up period, 301 days; IQR: 212-596 days). The atrial pacing leads were positioned in the right atrial (RA) septum in 17 patients (septal group) and in the RA appendage in 54 patients (RA appendage group), with significantly higher sLVP rates in the septal group compared with the RA appendage group (81% ± 30% vs 63% ± 37%; P = 0.045). In patients with first-degree atrioventricular blocks, the sLVP rates tended to be higher in the septal group. Therefore, RA septal pacing increased sLVP rates in patients undergoing CRT.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.