Abstract

This paper discusses the testing accomplished to determine the radio frequency (RF) susceptibility of a typical electroencephalograph (EEG) and to determine the effectiveness of lead X-ray shielding as an RF shield. New data are presented on threshold field strengths and frequencies which just begin to cause interference on a typical EEG brain scan. Results are combined with data from other studies to show that a typical EEG machine would be immune to the RF environment expected in the vicinity of most hospitals. Therefore, expensive RF shielding in a hospital EEG suite is usually not required. Finally, an inexpensive shielding system utilizing X-ray shielding which was developed by the U.S. Veterans Administration is presented.