Abstract
This study describes the MRI-related radio frequency (RF) safety evaluation of breast tissue expander devices to establish safety criteria. Numerical simulations and experimental measurements were performed at 64 MHz with a gel phantom containing a breast expander. Additionally, computational modeling was performed (64 and 128 MHz) with an adult female model, containing a virtually implanted breast tissue expander device for four imaging landmark positions. The presence of the breast tissue expander device led to significant alterations in specific absorption rate (SAR) and|B1+|distributions. The main source of SAR alterations with the use of the breast expander device was the saline-filled pouch of the expander. Conversely, the variation of RF magnetic field (B1+) was mainly caused by the metallic port. The measured values of electric field magnitude did not increase significantly due to the introduction of the expander device. The maximum 1g- or 10g-averaged SAR values in tissues near the implant were lower than those expected in other regions of the patient body with normalization of both|B1+|equal to 2 μT at the coil isocenter and whole body averaged SAR equal to 4W/kg.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE transactions on electromagnetic compatibility
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.