Abstract

With continuing hardware and pulse sequence advancements, modern MRI is gaining sensitivity to signals from short-T(2) (1)H species under practical experimental conditions. However, conventional MRI coils are typically not designed for this type of application, as they often contain proton-rich construction materials that may contribute confounding (1)H background signal during short-T(2) measurements. An example of this is shown herein. Separately, a loop-gap style coil was used to compare different coil construction materials and configurations with respect to observed (1)H background signal sizes in a small animal imaging system. Background signal sources were spatially identified and quantified in a number of different coil configurations. It was found that the type and placement of structural coil materials around the loop-gap resonator, as well as the coil's shielding configuration, are critical determinants of the coil's background signal size. Although this study employed a loop-gap resonator design, these findings are directly relevant to standard volume coils commonly used for MRI.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call