Abstract

A quadrupole inductively coupled plasma mass spectrometer (Q-ICP-MS) has been used for determination of lead in plant materials using isotope-dilution inductively coupled plasma mass spectrometry. The accuracy of the method was demonstrated by analysis of a matrix certified reference material, NIST SRM 1547 Peach Leaves. Specific instrumental parameters of Q-ICP-MS, including isotope analysis mode, integration time per point, number of points per mass, and number of measurements, were optimized to obtain the best measurement precision. The precision (expressed as relative standard deviation) associated with replicate measurement of the 208Pb/206Pb isotope ratio and its mass-bias correction factor was <0.2%. Following “Example A7” of the Eurachem/CITAC Guide, the relative expanded uncertainty, U rel, (coverage factor k = 2) was found to be ±1.1%, which fulfilled the target value of ±2% maximum and was lower than the uncertainty of ±3.4% reported by NIST based on isotope-dilution thermal ionization mass spectrometry. Sample recovery of 99% was obtained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.