Abstract

PurposeTo evaluate various selected Pentacam HR indices in a cohort of soft contact lens (SCL) wearers before and after SCL discontinuation.MethodsNinety right eyes of 90 SCL wearers (group 1) were compared to healthy controls (group 2). For group 1 eyes, Pentacam HR images were captured within one hour of SCL removal and after one month of discontinuation. The evaluated indices included curvature, elevation, and pachymetric parameters.ResultsElevation indices and two pachymetric indices (central corneal thickness and thinnest corneal thickness) did not show significant differences, either between the two groups or in group 1 after SCL discontinuation. Some pachymetric indices (average and maximum corneal pachymetry progression indices, and maximum Ambrosio’s relational thickness index) showed statistically significant (P-value <0.001), yet clinically irrelevant differences. Contrarily, changes in curvature indices were both statistically significant (P-value <0.001) and clinically relevant.ConclusionFrom a clinical point of view, many elevation and pachymetric indices are robust against SCL wear, and can be relied upon for corneal ectasia screening of SCL wearers, even without abstaining from wearing SCLs.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call