Abstract

Potentiometric membrane electrodes that respond to heparin and other polyanions were introduced in the early 1990s. Herein, the mechanism of polymer membrane electrode type heparin sensors is revisited. The extraction/diffusion of heparin is studied via both potentiometric and impedance spectroscopic techniques using a pre-fractionated heparin preparation that contains polyanionic species > 10000 Daltons. The reversal in EMF response using this heparin preparation indicates diffusion of higher MW heparin fragments to the backside of the membrane. Diffusion coefficients are calculated using a novel formula derived from the phase boundary potential model and Fick's second law of diffusion. Impedance spectroscopy is also employed to show that high MW heparin species are extracted and diffuse across the PVC membranes.