Abstract

Background Cochlear implant (CI) surgery is a safe surgical technique, although some patients require revision CI surgery. Aims/Objectives This study investigated the cause and underlying reason of revision CI surgery as well as hearing outcomes in a single institution. Patients and Methods This retrospective study evaluated patients who underwent CI surgery between April 2006 to March 2022 (n = 351). Sex, aetiology of hearing loss (HL), age and period from initial CI surgery to reimplantation, cause of revision, and related factors were examined. Results Twelve patients (8 males, 4 females) received CI reimplantation. The revision surgery rate was 2.59% (3.15% children, 1.69% adults); the period from initial surgery to reoperation was 8.60 ± 6.56 years for 9 children with congenital HL and 15.27 ± 5.72 years for 3 adults with progressive HL. Device failure was the most common cause (n = 8), followed by infections (n = 2), advanced facial irritation symptoms (n = 1), and electrode slip-out (n = 1). Mean preoperative and postoperative CI thresholds were 44.0 ± 9.46 dBnHL and 39.19 ± 8.89 dBnHL (p < .068), respectively. Conclusion and Significance Caregiver education, surgical technique advances, flap design, and extensive antibiotic use may decrease the revision surgery rate. The lack of post-revision deterioration of the hearing threshold contributed to well-being in patients with CI.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call