Abstract

Albedo systems are the mainstay at many facilities and continue to be refined. Advanced electrochemical etching techniques for CR-39 now yield a dose equivalent response that is nearly constant from 0.1 to 4.0 MeV. Recent studies include use of converters to enhance CR-39 response at both low and high energies. Methods have been suggested for use of CR-39, either alone or in conjunction with albedo and other detectors to provide spectral information as a step to more accurate dosimetry. Limitations in the use of CR-39 primarily center on the lack of consistent, high-quality, dosimetry-grade material, significant angular dependence, and poor dose equivalent response at both low and high energies. Work continues on silicon diodes, with some new designs. The most attractive new dosimetry technique is the bubble-damage or superheated drop detector. Metal-on-silicon (MOS) microelectronics present exciting possibilities for the future. 25 refs., 6 figs.