Abstract
Review is made of dosimetric studies of Ge-doped SiO2 telecommunication fibre as a 1-D thermoluminescence (TL) system for therapeutic applications. To-date, the response of these fibres has been investigated for UV sources, superficial X-ray beam therapy facilities, a synchrotron microbeam facility, electron linear accelerators, protons, neutrons and alpha particles, covering the energy range from a few eV to several MeV. Dosimetric characteristics include, reproducibility, fading, dose response, reciprocity between TL yield and dose-rate and energy dependence. The fibres produce a flat response to fixed photon and electron doses to within better than 3% of the mean TL distribution. Irradiated Ge-doped SiO2 optical fibres show limited signal fading, with an average loss of TL signal of ∼0.4% per day. In terms of dose response, Ge-doped SiO2 optical fibres have been shown to provide linearity to x and electron doses, from a fraction of 1Gy up to 2kGy. The dosimeters have also been used in measuring photoelectron generation from iodinated contrast media; TL yields being some 60% greater in the presence of iodine than in its absence. The review is accompanied by previously unpublished data.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.