Abstract

A reverse intensity correction method was developed for spectral library searches to correct for instrument response without the side effect of magnifying the noise in the low responsivity region of test spectra. Instead of applying relative intensity correction to the sample test spectra to match the standardized library spectra, a reverse intensity correction is applied to the standardized library spectra to match the uncorrected sample spectrum. This simple procedural change improves library search performance, especially for dispersive charge-coupled device Raman analyzers using near-infrared excitations, where the instrument response often varies greatly across the spectral range, and signal-to-noise ratio in the low responsivity regions is typically poor.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.