Abstract

The purpose of this study was to describe a modified technique for sutureless intrascleral intraocular lens fixation in in patients without capsular support: the reverse 4-flanged technique. A 2.2 mm corneal incision was made for aphakic patients. The 6-0 polypropylene sutures were threaded through a 30-gauge needle outside the eye. An MDJ® injector was employed to insert the IOL in the eye. A suture-needle snare with 7-0 polypropylene was used to exteriorize the superior end of the sutures through the sclerotomy. The 6-0 polypropylene was heated with a cautery to create the flanges. Nine eyes of nine patients underwent the reverse 4-flanged technique. The median follow-up time was 182 days (range 174 - 195). Best corrected distance visual acuity (BCVA) improved from 20/400 [hand movement - 20/33] to 20/120 [20/400-20/21]. No vitreous haemorrhage, retinal detachment, endophthalmitis, IOL luxation or flange exposure was recorded during the follow-up. The reverse 4-flanged technique maintains the advantages of the original 4-flanged technique in terms of IOL stability while reducing the incision size and surgical manoeuvres performed inside the eye. This modification promises to be a reliable, safe and relatively simple technique to correct aphakia in the absence of capsular support.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.