Abstract

Reliable securement of peripheral venous cannulae (PVC) is an important factor in their maintenance. This audit in a district general hospital compares the occurrence of PVC restarts between a 3-month period in 2010 and the same 3 months in 2011. The only difference in the PVC care bundle between these dates was the implementation of an advanced securement dressing for cannulae in 2011. Results show a significant increase in cannulae attaining the maximum local protocol duration of 72 hours during 2011. Also, restarts owing to dressing-influenced factors (dislodgement, infiltration and leakage) were significantly lower in 2011 when the new dressing was used. The total number of PVC restarts during the comparative audit periods was 9% lower in 2011 compared with 2010. This data suggests that better PVC securement is leading to an overall reduction in PVC insertions but further evidence is required to support this conclusion.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.