Abstract

In a Clinical Laboratory Improvement Amendments laboratory setting, we evaluated severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) IgG detection with 4 lateral flow immunoassays [LFIAs; 2 iterations from BTNX Inc. (n = 457) and 1 each from ACON Laboratories (n = 200) and SD BIOSENSOR (n = 155)]. In a cohort of primarily hospitalized, reverse-transcription polymerase chain reaction–confirmed coronavirus disease 2019 cases, sensitivity at ≥14 days from symptom onset was: BTNX kit 1, 95%; BTNX kit 2, 91%; ACON, 95%; and SD, 92%. All assays showed good concordance with the Abbott SARS-CoV-2 IgG assay at ≥14 days from symptom onset: BTNX kit 1, 99%; BTNX kit 2, 94%; ACON, 99%; and SD, 100%. Specificity, measured using specimens collected prior to SARS-CoV-2 circulation in the United States and “cross-reactivity challenge” specimens, was 98% for BTNX kit 1 and ACON and 100% for BTNX kit 2 and SD. These results suggest that LFIAs may provide adequate results for rapid detection of SARS-CoV-2.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.