Abstract

To evaluate the feasibility of motion correction in glutamate chemical exchange saturation transfer (GluCEST) imaging, using a rat model of epileptic seizure. Epileptic seizure was induced in six male Wistar rats by intraperitoneal injection of kainic acid (KA). CEST data were obtained using a 7.0T Bruker MRI scanner before and 3h after KA injection. Retrospective motion correction was performed in CEST images using a gradient-based motion correction (GradMC) algorithm. GluCEST signals in the hippocampal regions were quantitatively evaluated with and without motion correction. Calculated GluCEST signals differed significantly between the pre-KA injection group, regardless of motion-correction implementation, and the post-KA injection group with motion correction (3.662 ± 1.393 % / 3.726 ± 1.982 % for pre-KA injection group with/without motion correction vs. 6.996 ± 1.684 % for post-KA injection group with motion correction; all P < 0.05). Our results clearly show that GradMC can be used in CEST imaging for efficient correction of seizure-like motion. The GradMC can be further implemented in various CEST imaging techniques to increase the accuracy of analysis.